Series RH/RHL
Humidity/Temperature Transmitter
 
Calibration-Free,2%Or 3%Accuracy,Optional Display
   


Humidity,Temperature Transmitter

 

Demanding humidity/temperature applications require the Series RH/RHL Transmitter which offers high accuracy,long term stability,and reliable operation.The Series RH/RHL is designed for monitoring and controlling humidity or both humidity and temperature in building energy management systems,HVAC,commercial,residential,clean rooms,museums,climate chambers,and other space monitoring applications.

The Series RH/RHL is a two-wire transmitter with a 4-20 mA loop powered output or 0 to 10 VDC output.The state of the art sensor recovers from 100%saturation and is calibration-free.A variety of mounting configurations are available including wall mount,duct mount,and OSA (outside air)models.Select humidity monitoring or humidity and temperature models.

The combined humidity/temperature version (RHT)provides dual 4-20 mA or 0 to 10 VDC output signals to control both humidity and temperature with one sensor which reduces installation costs.The duct
mount version is also available with an optional alpha-numeric LCD to provide local indication of humidity and temperature simultaneously. The displayed temperature is field selectable for °F or °C.Monitor humidity in ducts,rooms,and outside air.
 
Specifications :
Relative Humidity Range: 0 to 100%RH.
Temperature Range:-40 to 140 °F (-40 to 60 °C).
Accuracy:(RH):±2%@10 -90%RH;(RHL):±3%@20-80% RH;±0.9 °F @72 °F (±0.3 °C @25 °C).
Temperature Limits:-40 to 140 °F (-40 to 60 °C).
Storage Temperature:-40 to 176 °F (-40 to 80 °C).
Compensated Temperature Range:-4 to 140 °F (-20 to 60 °C).
Power Requirements: 10-35 VDC.
Output Signal: 4-20 mA or 0-10 VDC,2 channels for humidity/temperature models (loop powered on RH current models).
Response Time: 5-15 seconds.
Electrical Connections: Screw terminal block.
Conduit Connection: Duct mount:1/2 ? NPS;;OSA:1/2 ? ((22.3 mm).
Drift: <1%RH/year.
RH Sensor: Capacitance polymer.
Temperature Sensor: Solid state band gap.
Housing Material: Wall Mount:ABS;Duct Mount:PBT;OSA: Polycarbonate.
Enclosure Rating: NEMA 4X (IP65)for OSA mount only.
Display: Optional 2-line alpha-numeric,8 characters/line for duct mount only.
Display Resolution: RH:0.1%;0.1 °F (0.1 °C).
Weight:Wall Mount: 0.5 lb (0.25 kg);Duct Mount:0.6 lb (0.3 kg); OSA:1 lb (0.45 kg).
Agency Approvals: CE.

Copyright © 2008. ALM .
English Arabic French Spanish